Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Goerz SE400: CAL ADJ /5520 DATE: 04-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 94 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # ABB GOERZ AG SE400 Service manual No.331 0073 03 date:06.89 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP Set the WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP This procedure was intended to check the performance 1.006 DISP charistics of the SE400 with an 884 0811 00 plug-in 1.006 DISP measurement unit, and a time unit. 1.007 DISP Set the REC/<0> switch to <0>. 1.007 DISP Set the V/mV switch to V. 1.007 DISP Set the RANGE knob to 1 (V). 1.007 DISP Set the CAL/VER switch to CAL. 1.007 DISP Set the x100% switch to 0. 1.007 DISP Set the TIME UNIT to 6cm/min 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to ABB 884 INPUT HI 1.008 DISP [32] 5520A NORMAL LO to ABB 884 INPUT LO 1.009 HEAD V - mV OFFSET 1.010 DISP START the chart feed. 1.011 DISP Set the V/mV switch to V. 1.011 DISP Turn the ZERO control to set the trace at EXACTLY 50%. 1.011 DISP Set the V/mV switch to mV. 1.012 OPBR Is the trace at EXACTLY 50% (same as V setting)? 1.013 JMPT 1.016 1.014 DISP Adjust RP2 (offset) to set the trace at 50%. 1.015 JMP 1.011 1.016 DISP Set the V/mV switch to V. 1.016 DISP Set the RANGE knob to 2 (V). 1.016 DISP Set the CAL/VER switch to CAL. 1.017 DISP Set the ZERO control to 0%. 1.018 DISP Set the REC/<0> switch to REC. 1.019 5520 2V S 2W 1.020 DISP Adjust RP3 (chip) to set trace at 100%. 1.021 DISP Set the x100% switch to -1. 1.022 DISP Adjust RP9 (zero suppr) to set trace at 0%. 1.023 5520 0V S 2W 1.024 DISP Set the REC/<0> switch to <0>. 1.025 DISP Adjust RP4 (LHstop) to set trace at -0.5%. 1.026 DISP Set the REC/<0> switch to REC. 1.027 DISP Set the RANGE knob to 1 (V). 1.027 DISP Set the x100% switch to 0. 1.028 5520 1.1V S 2W 1.029 DISP Adjust RP6 (RHstop) to set trace at 100.5%. 1.030 5520 1mV S 2W 1.031 DISP Set the UUT V/mV switch to mV. 1.031 DISP Set the RANGE knob to 1 (mV) 1.032 DISP Adjust RP1 (chip 1mV) to set trace at 100%. 1.033 5520 * S 1.034 DISP Set the UUT V/mV switch to V. 1.034 DISP Set the RANGE knob to 1 (V). 1.034 DISP Set the TIME UNIT to 60cm/min. 1.035 5520 0.95Vpp 0.1H SQ S 2W 1.036 DISP Adjust RP5 (damping) for square corner. 1.037 END